Technique Specification
ECG
Lead Mode : 5 Leads ( I , ll , Ill , AVR , AVL , AVF , V )
Gain : 2.5mm / mV, 5.0mm /mV , 10mm /Mv, 20mm /mV
Heart Rate : 15-300 BPM ( Adult ) ; 15-350 BPM ( Neonatal )
Resolution : 1 BPM
Accuracy : ± 1%
Sensitivity > 200 uV ( Peak to peak )
ST measurement range : -2.0 ~+ 2.0 Mv
Accuracy : -0. 8mV~+ 0. 8mV : ± 0. 02mV or ± 10% , which is greater
Other Range : unspecified
Sweep speed : 12.5 mm / s , 25mm / s , 50mm / s
Bandwidth:
Diagnostic : 0.05~130 Hz
Monitor : 0.5~40 Hz
Surgery : 1~20 Hz
SPO2
Measuring Range : 0 ~ 100%
Resolution : 1%
Accuracy : 70%~100% ( ± 2 % )
Pulse Rate : 20-250 BPM
Resolution : 1 BPM
Accuracy : ± 3 BPM
Optinal Parameters :
Central monitoring system
Dual IBP
Etco 2
12-lead ECG
Nellcor SpO2
Sun Tech NIBP
NIBP
Method : Oscillometry
Measure mode Manual, Auto, STAT
Unit : mmHg , kPa
Measure and alarm range:
Adult Mode
SYS 40 ~ 280 mmHg
DIA 10 ~ 225 mmHg
MEAN 20 ~ 240 mmHg
Pediatric Mode
SYS 40 ~ 220 mmHg
DIA 10 ~ 160 mmHg
MEAN 20 ~ 170 mmHg
Neonatal Mode
SYS 40 ~ 135 mmHg
DIA 10 ~ 100 mmHg
MEAN 20 ~ 110 mmHg
Resolution : 1 mmHg
Accuracy : ± 5 mmHg
TEMP
Measure and Alarm Range : 0 ~ 50 C
Resolution : 0 . 1C
Accuracy : ± 0 . 1C
Standard Parameters
ECG, RESP, TEMP, NIBP, SPO2, PR
RESP
Method : Impedance between RA-LL or RALA
Measurement Range:
Adult : 10~120 BrPM
Neonatal / Pediatric : 7~150 BrPM
Resolution : 1 BrPM
Accuracy : ± 2 BrPM
Standard Configuration